HysteroFlow/Hystero Balance II (HFHB-2)

Fluid Management

HysteroFlow/Hystero Balance II (HFHB-2)

The Olympus HysteroFlow/HysteroBalance II Fluid Management System provides accurate and easy measurement of fluid loss and intrauterine pressure for a variety of diagnostic and therapeutic procedures.


Key Benefits

  • User-friendly display: The system features a large touchscreen display monitor with a text-based user menu and simple navigation.
  • Robust system cart: The system cart includes two weighing scales and can hold up to four fluid irrigation bags and six collection canisters.
  • Simple operation: The system provides easy bag and container changes without the need to pause a running procedure.

Product Support

WA40634A‡ Tubing set for 2 irrigation bags sterile, inflow and outflow, 5/pkg
WA40635A‡ Tubing set for 2 irrigation bags, sterile, inflow only, 10/pkg
006775-901‡ Under Buttocks Drape/Fluid Collection Set, 5/pkg
006772-901 3L Canisters, 36/pkg

‡ Supplied sterile, single-use, disposable

Main Connections

Voltage 100–240 V
Frequency 50–60 Hz
Power consumption (max.) 170 VA

Safety

Protection class I
Application part type (B, BF, CF) BF
Impact protection IP41 (WA40620A), IP21 (WA40622A)
Classification IIb

Environment

Operating conditions 10 to 40 °C
30–75 % rel. humidity
70 to 106 kPa air pressure
3000 m max. altitude above sea level
for device use
Used with flammable anesthetic gases This device is not
designed for use with flammable
anesthetic agents (class APG).
Storage and transportation conditions -20 to 70 °C
10 to 90 % rel. humidity
70 to 106 kPa air pressure

Pump
Adjustable Values

Flow 30 to 500 ml/min
Pressure 35 to 150 mmHg

Measurement Range

Flow 0 to 500 ml/min
Pressure 0 to 750 mmHg
Deficit ±22000 ml

Dimensions
Size (Width x Height x Depth)

WA40620A 370 x 157 x 396 mm
WA40622A 640 x 1145 x 655 mm

Weight

WA40620A 6.1 kg
WA40622A 30 kg
Max. load 2 x 22 kg

Olympus Service & Repair

Olympus offers a broad range of services to healthcare professionals and to our customers, including contact hour and peer-based training courses; information, training tools and videos on infection control and reprocessing; authorized repair services and support on Olympus equipment; and financing solutions to help your facility with acquisition of new capital equipment, accessories, and maintenance plans.

Need Help?

Cleaning, Disinfection & Sterilization

The proper cleaning, disinfection, and sterilization of Olympus equipment is equally as important as their proper use. 

View Products Information Instructional Videos
OER-Pro Infection Prevention Bronchoscope Reprocessing
All Products Resources GI Endoscope Reprocessing
    Ultrasound Reprocessing

 

Olympus University

Olympus Professional Education also offers training programs through Olympus University (OU). Geared toward nurses, reprocessing technicians and biomeds, many of the OU courses are available for nursing contact hour continuing education credit.

Visit Olympus University